The HPLC chromatogram is the picture behind the purity number on a peptide COA. Most readers skip straight to the percentage, but the trace itself shows how that percentage was calculated and whether anything looks unusual. This guide walks through the figure. For the underlying techniques see HPLC and mass spectrometry explained.
The Axes
The horizontal axis is time, usually in minutes from the start of the run. The vertical axis is detector response, typically UV absorbance in milli-absorbance units (mAU). Peptide methods commonly detect at 210 to 220 nm, where the peptide bond itself absorbs, so nearly every peptide-related species in the sample shows up.
Retention Time
Each peak appears at a characteristic retention time. In reversed-phase HPLC, the standard method for peptides, a C18 column holds molecules by hydrophobicity, and a gradient of increasing organic solvent (usually acetonitrile with a small amount of TFA) releases them in order. More hydrophobic species elute later. Retention time on its own does not identify a compound; that is the job of mass spectrometry.
The Main Peak and the Small Ones
In a high-purity sample, one tall peak dominates. Smaller peaks nearby are usually synthesis-related impurities: deletion sequences missing one residue, truncated chains, incompletely deprotected species, or oxidized and deamidated forms. Peaks very close to the main peak are the hardest to separate and are the ones a good method is designed to resolve. See peptide synthesis methods and degradation pathways.
The Peak Table
Below the trace, most reports print a table listing each integrated peak with its retention time, area and area percent. The purity value is the main peak's area as a percentage of the total integrated area. If the main peak is 99.37% of the area, the remaining 0.63% is spread across the minor peaks. Checking that the table's main-peak percentage matches the headline purity is a quick consistency test.
Baseline and Integration
Integration software draws a baseline under each peak. A flat, quiet baseline makes small impurities easy to see. A noisy or drifting baseline can hide them, and aggressive integration settings can leave small peaks uncounted, inflating the purity number. Experienced readers look at whether the visible minor peaks appear in the table.
What HPLC Purity Does Not Measure
- Identity: a single clean peak could be the wrong molecule. MS confirms identity.
- Net peptide content: water, counter-ions and salts are not UV-active at the same level and are not counted. See net peptide content vs purity and acetate vs TFA salt forms.
- Endotoxin or microbial content: separate assays. See endotoxin testing.
Blends Show More Than One Main Peak
A two-peptide blend should show two major peaks, each integrated separately. See the Tesamorelin + Ipamorelin blend overview for an example of what a two-component report should include.
See Real Examples
The certificates in our COA library include the HPLC traces for published lots. Try reading one with this guide in hand: find the main peak, read its retention time, and check the peak table against the headline purity. Reports describe tested samples; confirm the compound, size and lot supplied before relying on any report.
Related Reading
how to read a peptide COA, how to verify a COA, COA glossary, and the pillar Researcher's Guide to Peptides in 2026.
Frequently Asked Questions
What does retention time mean on a peptide chromatogram?
It is the time from the start of the run at which a species leaves the column and reaches the detector. In reversed-phase HPLC, more hydrophobic molecules generally elute later.
How is HPLC purity calculated?
It is the area of the main peak as a percentage of the total integrated peak area, measured with UV detection, typically at 210 to 220 nm for peptides.
What are the small peaks next to the main peak?
Usually synthesis-related or degradation-related impurities such as deletion sequences, truncated chains, or oxidized and deamidated forms.
Does a clean HPLC trace confirm the peptide identity?
No. HPLC shows how much of the sample is one main component. Mass spectrometry is needed to confirm that the component has the expected molecular mass.
Why might HPLC purity differ from net peptide content?
HPLC purity compares peptide species with each other, while net peptide content also accounts for water, counter-ions and salts in the powder.
All products referenced in this article are sold strictly for laboratory research use only. They are not for human or veterinary use, are not drugs, foods, or cosmetics, and are not intended to diagnose, treat, cure, or prevent any disease. A Certificate of Analysis is an analytical record for a tested sample; it is research verification only and is not a statement of safety or suitability for any use outside a qualified laboratory.